Dental Testing Requirements
There are various aspects of testing that are required under the H.A.R.P. ACT.
A number of forms are required by the Ontario Ministry of Health and Long Term Care when a new office is opened, new equipment is purchased or there is a change of ownership. Government Forms
Dental Testing Requirements
Items 2,3 and 4 from Table 3 of the HARP Act (reproduced below) outline the annual testing requirements for dental x-ray machines.
DENTAL FACILITY
| Item | Column 1 |
Column 2 |
| Test or Procedure | Frequency | |
| 1. | Photographic quality control* | Every operational day |
| 2. | Patient entrance exposure measurements | Every twelve months and upon alteration or servicing of the machine |
| 3. | Collimation | Every twelve months and upon alteration of servicing of the machine |
| 4. | Half-value layer | Every twelve months and upon alteration or servicing of the machine |
R.R.O. 1990, Reg. 543, Table 3.
* For dental offices using digital radiography the requirement for daily photographic quality control has been waived.
Varanidex provides the annual QC testing for dental x-ray machines either by its convenient, cost effective mail service, or by having its personnel perform the testing at the dental office. More than 80% of our clients use our mail service. For further information on this service please contact us by phone or e-mail.
Acceptance Testing is required when a new (or used) machine is installed, or when a machine is moved from one location to another. This is a bit more robust than the annual requirement. Varanidex staff perform these tests at the dental office.