Dental Testing Requirements

There are various aspects of testing that are required under the H.A.R.P. ACT.

A number of forms are required by the Ontario Ministry of Health and Long Term Care when a new office is opened, new equipment is purchased or there is a change of ownership. Government Forms

Dental Testing Requirements

Items 2,3 and 4 from Table 3 of the HARP Act (reproduced below) outline the annual testing requirements for dental x-ray machines.

DENTAL FACILITY

Item Column 1

Column 2

Test or Procedure Frequency
1. Photographic quality control* Every operational day
2. Patient entrance exposure measurements Every twelve months and upon alteration or servicing of the machine
3. Collimation Every twelve months and upon alteration of servicing of the machine
4. Half-value layer Every twelve months and upon alteration or servicing of the machine

R.R.O. 1990, Reg. 543, Table 3.

* For dental offices using digital radiography the requirement for daily photographic quality control has been waived.

Varanidex provides the annual QC testing for dental x-ray machines either by its convenient, cost effective mail service, or by having its personnel perform the testing at the dental office. More than 80% of our clients use our mail service. For further information on this service please contact us by phone or e-mail.

Acceptance Testing is required when a new (or used) machine is installed, or when a machine is moved from one location to another. This is a bit more robust than the annual requirement. Varanidex staff perform these tests at the dental office.

For a quote for testing at your office, or for assistance in aspects of the HARP Act – please do not hesitate to contact our office.